Technical data Smart
| TECHNICAL DATA |
|
| air convection |
forced / natural |
| chamber capacity [l] |
200 / 85 |
| working capacity [l] |
163 / 73 |
| controller |
microprocessor PID |
| display |
4,3″ full colour touch screen |
| TEMPERATURE |
|
| temperature range [°C] |
0…+15 / -25…0 |
| temperature resolution every … [°C] |
0,1 |
| temperature fluctuation at 4°C [+/-°C]* |
0,4 |
| temperature fluctuation at -20°C [+/-°C]* |
0,5 |
| temperature variation at 4°C [+/-°C]* |
0,7 |
| temperature variation at -20°C [+/-°C]* |
2 |
| temperature protection |
class 1.0 to DIN 12880 / class 3.2 (option) |
| CHAMBER |
|
| door type |
solid / glass or double (option) /4/ | solid (ZL) |
| interior |
|
| C Smart |
stainless steel to DIN 1.4016 |
| CS Smart |
stainless steel to DIN 1.4016 |
| P Smart |
acid-proof stainless steel to DIN 1.4301 |
| PS Smart |
acid-proof stainless steel to DIN 1.4301 |
| housing |
|
| C Smart |
powder coated sheet |
| CS Smart |
stainless steel polished |
| P Smart |
powder coated sheet |
| PS Smart |
stainless steel polished |
| overall dims [mm] /1/ |
|
| width A |
610 |
| height B |
1940 |
| depth C |
650 |
| internal dims [mm] |
|
| width D |
480 / 400 |
| width D’ |
520 / 420 |
| height E |
660 / 590 |
| depth F |
420 / 400 |
| depth F’ |
480 / 440 |
| depth G |
320 / 270 |
| height H |
640 / 380 |
 |
| shelves (standard | max) |
3 | 4 / 2 | 4 |
| max shelf workload [kg] /2/ |
10 |
| max unit workload [kg] |
30 / 30 |
| weight [kg] |
121 |
| ELECTRICAL PARAMETERS |
|
| voltage |
230V 50-60Hz |
| nominal power [W] |
500 |
| refrigerant |
R1234ze / GWP=7 | R290 / GWP=3 |
| warranty |
24 months |
| manufacturer |
POL-EKO® |
all the above technical data refer to standard units (without optional accessories)
* – fluctuation measured in centre of the chamber; in space, variation (K) calculated for chamber as: K= +/- (T average max. – T average min. ) / 2; parameters given for the chamber above the bottom step
** – other power supplies on request
1 – depth doesn’t include 50 mm of power cable, the width does not include the 20 mm of rubber plug
2 – on uniformly loaded surface
3 – reinforced shelf
4 – additional internal glass door
Technical data Smart PRO
| TECHNICAL DATA |
|
| air convection |
forced / natural |
| chamber capacity [l] |
200 / 85 |
| working capacity [l] |
163 / 73 |
| controller |
microprocessor PID |
| display |
7″ full colour touch screen |
| TEMPERATURE |
|
| temperature range [°C] |
0…+15 / -25…0 |
| temperature resolution every … [°C] |
0,1 |
| temperature fluctuation at 4°C [+/-°C]* |
0,4 |
| temperature fluctuation at -20°C [+/-°C]* |
0,5 |
| temperature variation at 4°C [+/-°C]* |
0,7 |
| temperature variation at -20°C [+/-°C]* |
2 |
| temperature protection |
class 3.2 to DIN 12880 |
| CHAMBER |
|
| door type |
solid / glass or double (option) /4/ | solid (ZL) |
| interior |
|
| C Smart PRO |
stainless steel to DIN 1.4016 |
| CS Smart PRO |
stainless steel to DIN 1.4016 |
| P Smart PRO |
acid-proof stainless steel to DIN 1.4301 |
| PS Smart PRO |
acid-proof stainless steel to DIN 1.4301 |
| housing |
|
| C Smart PRO |
powder coated sheet |
| CS Smart PRO |
stainless steel polished |
| P Smart PRO |
powder coated sheet |
| PS Smart PRO |
stainless steel polished |
| overall dims [mm] /1/ |
|
| width A |
610 |
| height B |
1940 |
| depth C |
650 |
| internal dims [mm] |
|
| width D |
480 / 400 |
| width D’ |
520 / 420 |
| height E |
660 / 590 |
| depth F |
420 / 400 |
| depth F’ |
480 / 440 |
| depth G |
320 / 270 |
| height H |
640 / 380 |
 |
| shelves (standard | max) |
3 | 4 / 2 | 4 |
| max shelf workload [kg] /2/ |
10 |
| max unit workload [kg] |
30 / 30 |
| weight [kg] |
121 |
| ELECTRICAL PARAMETERS |
|
| voltage |
230V 50-60Hz |
| nominal power [W] |
500 |
| refrigerant |
R1234ze / GWP=7 | R290 / GWP=3 |
| warranty |
24 months |
| manufacturer |
POL-EKO® |
all the above technical data refer to standard units (without optional accessories)
* – fluctuation measured in centre of the chamber; in space, variation (K) calculated for chamber as: K= +/- (T average max. – T average min. ) / 2; parameters given for the chamber above the bottom step
** – other power supplies on request
1 – depth doesn’t include 50 mm of power cable, the width does not include the 20 mm of rubber plug
2 – on uniformly loaded surface
3 – reinforced shelf
4 – additional internal glass door